QA
- Oversees the quality system.
- Reviews evidence, including the batch record, the event log, and the audit trail.
- Approves procedures and records.
- Makes the release decision. In this scenario the disposition decision is human.
Jonathan Muteba — Life Sciences Capability Showcase
Synthetic capability demonstration.
ROLE
QC tests the product. QA oversees the system, reviews evidence, approves procedures and records, and makes release decisions.
FDA regulations say quality control unit. ICH Q7 says quality unit. QA does not execute the chromatography step or the QC tests.
SCENARIO
In this scenario QA reviews CHR-401 from qualification through the batch on the skid.
The CHR-401 OQ report arrived with protocol discrepancy PD-CHR-004 still open. QA reviews the report against the executed protocol and does not approve it while the discrepancy is open.
The report due FridayA cracked UV flow cell on CHR-401 was replaced with PN-UV-22 instead of the approved PN-UV-20. The window material differs. Return to GMP use waits for QA approval.
Same part, different partThe paper record for DS-204-26-007 says N for unplanned holds. The CHR-401 event log shows a 14-minute hold. The record does not match the event log.
The hold nobody wrote downPATH
In this scenario a requirement becomes a record in four steps.
1
21 CFR 211.22 assigns the quality control unit responsibility for approving or rejecting components, containers, closures, in-process materials, packaging, labeling, and drug products. In this scenario QA oversees, and a person makes the disposition decision.
2
ICH Q7 is the GMP guide for this drug substance. Section 6.71 calls for the quality unit to review the batch production record of a critical process step before release. FDA data integrity guidance names attributable, legible, contemporaneous, original, and accurate.
3
Per site procedure, a manual hold is written on the batch record. A late entry is labeled late and does not overwrite the original line.
4
The paper batch record for DS-204-26-007 says N. The CHR-401 event log shows a 14-minute hold. The record does not match the event log, and DEV-26-0412 is open.
RECORD
FDA's data integrity guidance names five ALCOA attributes. The four plus attributes are a widely used industry extension.
Synthetic logbook entry for CHR-401, post-load wash, cycle 2 of 3.
14:22 OP-204 CHR-401 cycle 2 of 3, post-load wash. Pressure warning, peak 2.6 bar. Manual hold started to check inlet filter. Hold ended 14:36 (14 min). Entered 14:36.
Attributable
Named in FDA's data integrity guidance
OP-204 identifies the person who wrote this line.
Legible
Named in FDA's data integrity guidance
The wash step, the pressure warning, and the hold can be read as written.
Contemporaneous
Named in FDA's data integrity guidance
Written at the time of the event: 14:22 to 14:36.
Original
Named in FDA's data integrity guidance
This line is the original entry for the hold. It records the wash as it happened.
Accurate
Named in FDA's data integrity guidance
Peak 2.6 bar and the 14-minute hold match the CHR-401 event log.
Complete
Widely used industry extension
The warning, the inlet-filter check, and the hold from 14:22 to 14:36 are on the line.
Consistent
Widely used industry extension
The line agrees with the CHR-401 event log. In this scenario the paper batch record says N, so the two records do not match.
Enduring
Widely used industry extension
The line is kept on the controlled logbook, not a loose note.
Available
Widely used industry extension
A reviewer can retrieve the line with the event log for DS-204-26-007.
If an entry is missed, a late entry is labeled late, dated when written, gives a reason, and never overwrites the original. In this scenario OP-204 adds that late entry at step 7.4 on Day 3 for DS-204-26-007, under DEV-26-0412, after an operator statement. The original N stays visible.
The EU revised Chapter 4 is a draft. The 2025 revisions of Annex 11 and Chapter 4 are drafts.
The 2011 versions of EU GMP Annex 11 and Chapter 4 are the versions in force.
ALCOA+SOURCES
Each name links to the official source. 21 CFR 211 is the general teaching reference on this page. The CHR-401 case cites ICH Q7.
Document
What it covers
How this showcase uses it
Responsibilities of the quality control unit.
Teaching reference for oversight. QC tests the product. QA oversees, reviews, and approves.
Production record review.
General teaching reference for batch record review. The CHR-401 case cites ICH Q7 for the drug substance.
Good manufacturing practice guide for active pharmaceutical ingredients, including 2.22, 6.7, 6.53, and 6.71.
Chromatography in this scenario makes drug substance. The quality unit reviews the batch record, and the investigation looks at other batches that may be associated with the deviation.
Quality risk management.
Classification, root cause, and disposition stay with a person. The queue score does not make the decision.
A model for a pharmaceutical quality system across the product lifecycle.
The CHR-401 path in this scenario runs from qualification, through manufacturing, to batch review.
Data integrity and compliance with drug CGMP. December 2018.
Names the five ALCOA attributes on this page. The four plus attributes are not credited to this guidance.
Computerised systems. The linked text is the January 2011 version.
In this scenario the reviewer compares the CHR-401 event log and audit trail with the paper batch record. This page does not claim Annex 11 compliance.
Documentation. The linked text is the January 2011 version.
The revised Chapter 4 is a draft. This page uses the 2011 documentation chapter as the teaching reference for how a record is kept.
Share of pharma staff time spent on documentation-related activities.
about 30 percent
Activity time for one minor deviation.
about 18.1 hours
Calendar time to complete a minor deviation.
about 29 days
The queue and the disposition gates are on the operations center. The five cases and the ALCOA+ drill are in the Decision Lab.